Technical DATA
Description of the unit
|
Classification EN 60601-1 |
Class I
Applied part, Type B
IP 20, unit
IP X8, electrical foot pedal
Mode: continuous operation |
|
Classification 93 / 42 EC |
Class IIb |
|
Supply voltage |
100 – 127 – 220 – 240 VAC |
|
Mains frequency |
50 – 60 Hz |
|
Power consumption |
150 VA |
|
Noise level |
65 dB(A) |
|
Fuses |
T 3.15 L |
|
Operating mode |
Multiple pulse mode
repetition rate: 1 – 20 Hz |
|
Weight |
approx. 14.5 kg (approx. 30.8 lbs) |
|
Dimensions in mm (LxWxH) |
395 mm x 400 mm x 170 mm |
|
Operating conditions |
+10°C to +35°C
30% to 75% relative humidity
atmospheric pressure 700 hPa to 1060 hPa |
|
Storage and transport conditions |
+5°C to +40°C
10% to 90% relative humidity
700 hPa to 1060 hPa air pressure
Atmospheric pressure 700 hPa to 1060 hPa |
|
|
Description of the handpiece
|
Classification EN 60601-1 |
Class I
Applied part, Type B
IP 20 |
|
Aperture angle |
94° |
|
Aperture diameter |
100 mm |
|
Energy flux density (positive) |
0.03 to 0.4 mJ/mm2 |
|
Pressure range |
11.5 to 82.2 MPa |
Focal size (min. to max.):
fx(-6dB) = fy(-6dB)
fz(-6dB) |
1.2 to 2.8 mm
5.0 to 14.1 mm
|
|
Focal volume |
58 mm3 |
|
Pulse frequency |
1 to 8 Hz |
|
Intensity stages |
20 steps |
|
Overall energy E (5 mm) |
9.54 mJ |
|
Penetration depth |
5 to 40 mm |
|
Weight |
approx. 1.2 kg (approx. 2.7 lbs) |
|
Operating conditions |
+10°C to +35°C
30% to 75% relative humidity
atmospheric pressure 700 hPa to 1060 hPa |
|
Storage and transport conditions |
+5°C to +40°C
10% to 90% relative humidity
700 hPa to 1060 hPa air pressure
atmospheric pressure 700 hPa to 1060 hPa |
|
|